High-speed 2D Optical Micrometer

TM-3000 series

Catalogs View Catalog

Specs High-speed 2D Optical Micrometer TM-3000 series

Model

TM-006

TM-040

TM-065

Image

Measuring range

ø6 mm 0.24"

ø40 mm 1.57"

ø65 mm 2.56"

Minimum detectable object

0.04 mm 0.001"

0.3 mm 0.01"

0.5 mm 0.02"

Transmitter/receiver distance

60 mm 2.36"

180 mm 7.09"

270 mm 10.63"

Light source

GaN Green LED

InGaN Green LED

Measurement accuracy

±0.5 µm ±0.000020" *1

±2 µm 0.000079" *5

±3 µm ±0.000119"*7

Repeatability

±0.06 µm*2

±0.15 µm ±0.000006" *6

±0.2 µm 0.000008" *8

Sampling cycle (trigger interval)

5.5ms (33ms at the initial setting)*3

Environmental resistance

Enclosure rating

IP64*4

Ambient temperature

0 to +50 °C 32 to 122 °F

Relative humidity

35 to 85 % RH (No condensation)

Material

Aluminum

Weight

Transmitter

Approx 140 g

Approx 560 g

Approx 1280 g

Receiver

Approx 340 g

Approx 720 g

Approx 1460 g

Base

Approx 220 g

Approx 630 g

Approx 1500 g

*1 In a measurement area of 2 mm 0.08"× ø4 mm ø0.16" error when measuring width of KEYENCE standard object (glass calibration scale).
*2 Value of ±2σ measuring the width of KEYENCE standard object (glass calibration scale) in the centre of the measurement area, an average 16 times, average 1.3 mm 0.05" line.
*3 When measurement area is minimum, others are initial settings
*4 Apart from connector component
*5 In a measurement area of 10 mm 0.39"× ø26 mm ø1.02" error when measuring width of KEYENCE standard object (glass calibration scale).
*6 Value of ±2σ measuring the width of KEYENCE standard object (glass calibration scale) in the center of the measurement area, an average 16 times, average 8 mm 0.31" line.
*7 Error when measuring width of KEYENCE standard object (glass calibration scale) in a measurement area of 20 mm 0.79"× ø40 mm ø1.57".
*8 Value of ±2σ measuring the width of KEYENCE standard object (glass calibration scale) in the center of the measurement area, an average 16 times, average 14 mm 0.55" line.

Back to top

Model

TM-3001

TM-3001P

Image

Head compatibility

Compatible

Number of connectable sensors

2 units max.*1

Display

Minimum display unit

0.01 µm, 0.001 mm2, 0.01 °

Maximum display range

±9999.99 mm ±393.7", ±99999.9 mm2, ±99999.9°

Input terminal block

Laser remote interlock input

Non-voltage input*2

Trigger input (for Head A)

Non-voltage input*2

Voltage input*2

Timing 1 input

Auto-zero 1 input

Reset input

Output terminal block

Analog voltage output

±10 V x 2 outputs, out put impedance: 100 Ω*2

Total judgment output

NPN open-collector output*2

PNP open-collector output*2

Error output

NPN open-collector output (N.C.)*2

PNP open-collector output (N.C.)*2

Process output

NPN open-collector output*2

PNP open-collector output*2

Trigger input enable output

Adjusted error output

Expansion connector

Trigger input (for Head A)

Non-voltage input*2

Voltage input*2

Timing 2 input

Auto-zero 2 input

Program switching input

Non-voltage input, 4 inputs*2

Voltage input, 4 inputs*2

Memory card save input

Non-voltage input*2

Voltage input*2

Judgment/Binary output

3-level judgment output: OUT1 to OUT16, total judgment output
Binary output: OUT1 to OUT16 measured data output (21 bits)
NPN open-collector output*3

3-level judgment output: OUT1 to OUT16, total judgment output
Binary output: OUT1 to OUT16 measured data output (21 bits)
PNP open-collector output*3

Strobe output

NPN open-collector output*2

PNP open-collector output*2

Trigger input enable output

Analog RGB monitor output

SVGA (800 x 600 pixels)

RS-232C interface

Measured data output and control input/output (Maximum baud rate: 115200 bps, selectable)

USB interface

In conformity with USB Revision 2.0 HI-SPEED (USB 1.1 Full-SPEED compatible)

Ethernet interface

1000BASE-T/1000 BASE-TX/10 BASE-T

Memory card

SD card CA-SD4G (4GB), CA-SD1G (1GB) support

Rating

Power voltage

24 VDC ±10 %, Ripple (P-P) 10 % or less

Current consumption.

1 head connected 480mA max./ 2 heads connected 550mA max.

Environmental resistance

Ambient temperature

0 to +50 °C 32 to 122 °F

Relative humidity

35 to 85 % RH (No condensation)

Material

Polycarbonate

Weight

Approx. 1,120 g

*1 1 or 2 units can be connected only with the same head model
*2 • The rating of the NPN/PNP open collector output (output terminal block): 50 mA (30 V or less) max., residual voltage: 1.4 V or less (50 mA) 1.0 V (20 mA)
• The rating of the NPN/PNP open collector output (expansion connector): 50 mA (30 V or less) max., residual voltage: 1.0 V or less
• Rating for non-voltage input, ON voltage 1V max., OFF current 0.3mA max. (trigger input terminal, ON voltage 5V max., OFF current 1mA max.)
• Voltage rating, maximum rating 26.4V, ON voltage 10.8V, OFF current 0.3mA (trigger input terminal maximum rating 26.4V, ON voltage 10.8V, OFF current 1mA)
*3 • The rating of the NPN/PNP open collector output (output terminal block): 50 mA (30 V or less) max., residual voltage: 1.4 V or less (50 mA) 1.0 V (20 mA)
• The rating of the NPN/PNP open collector output (expansion connector): 50 mA (30 V or less) max., residual voltage: 1.0 V or less
• Rating for non-voltage input, ON voltage 1V max., OFF current 0.3mA max. (trigger input terminal, ON voltage 5V max., OFF current 1mA max.)
• Voltage rating, maximum rating 26.4V, ON voltage 10.8V, OFF current 0.3mA (trigger input terminal maximum rating 26.4V, ON voltage 10.8V, OFF current 1mA). OUT 1 to OUT 8 decision result, OUT 9 to OUT 16 decision result, time share output of binary measurement data.